Chapter 8: Nutrition and fluid balance
Skip chapter table of contents and go to main content
PEG, percutaneous endoscopically placed gastrostomy.
Percutaneous endoscopically placed gastrostomy (PEG) placement and care
Pre‐procedural considerations
Prior to performing PEG care the patient's medical and nursing notes should be consulted to identify the tube placement date and method, and any infection control issues. The endoscopist's instructions should be noted, including any specific detail with regard to post‐placement care, for example suture removal, tube rotation, checking the volume of the balloon (if appropriate) and dressing changes.
Procedure guideline 8.12
Percutaneous endoscopically placed gastrostomy (PEG) tube care
Table 8.15 Prevention and resolution (Procedure guideline 8.12)
Problem | Cause | Prevention | Action |
---|---|---|---|
Displaced PEG tube | Excessive tension being applied to the tube, either when moving the external fixator or by over‐tightening | Support the tube beneath the fixator when moving. | Liaise with medical team. |
Tube cracked | Prolonged use of clamp without repositioning | Regular tube care and review. Change position of clamp regularly. | Liaise with medical team to arrange replacement. |
Tube blockage | Inadequate tube flushing or overuse for administering medications | Regular tube flushing. Avoid/minimize use of tube for medications. Use appropriate medication formulation. | See Procedure guideline 8.20: Enteral feeding tubes: unblocking. |
Procedure guideline 8.13
Radiologically inserted gastrostomy (RIG) tube care
Procedure guideline 8.14
Removal of T‐fasteners
The integrity of the water‐filled balloon is important in ensuring that the balloon gastrostomy remains in place. This is usually done 1 week post‐insertion and on a weekly basis while the balloon gastrostomy tube is in situ. The balloon is deflated and reinflated using an ENFit syringe via the balloon inflation valve.
Procedure guideline 8.15
Changing the balloon water and checking the volume of a balloon gastrostomy
Procedure guideline 8.16
Jejunostomy feeding tube care including dressing change
Local policy and guidelines (where available) should be identified and followed when managing the usage of these tubes.
Post‐procedural considerations
Immediate care
Local protocols should specify the observations to be taken in the immediate post‐operative recovery period. These should include observation of blood pressure, heart rate, respiration rate and pain score in addition to checks of the stoma site for bleeding, leakage of gastric contents or tube displacement (NHSI [116]). If the patient is complaining of pain or there are signs of peritonitis or tube displacement, contact the medical team immediately. Peritonitis manifests as abdominal pain, fever and a raised white cell count in the initial days after placement. Transient external leakage of the stomach contents from the puncture canal can also suggest that internal leakage is occurring (NHSI [116]).
If there is pain on feeding, prolonged or severe pain post‐procedure or fresh bleeding, or external leakage of gastric contents then the feed should be stopped immediately. The patient should be reviewed by a senior clinician and the following considered: computed tomography (CT) scan, contrast study or surgical review (NHSI [116]).
Additional complications that may occur in the immediate period after gastrostomy insertion include aspiration pneumonia, haemorrhage and wound infection (Taheri et al. [177]). Colonic perforation is rare but may occur during the procedure and is likely to lead to peritonitis.
All assessments and care given should be clearly documented, including details of any observations noted and dressings used. The timing of future assessments should be planned to ensure that relevant aspects of care are followed up; for example, if a swab is taken to ascertain whether there is an infection, this needs to be followed up in order to act on the results of microbiology culture.
The patient should be advised not to use moisturizing creams or talcum powder around the stoma site. The grease in creams can cause the external retention device to slip, allowing movement of the tube and increasing the risk of hypergranulation, leakage and infection (Haywood [65]). Creams and talcs can affect the tube material, causing it to stretch or leak.
Once the stoma site has healed (approximately 10 days post‐insertion), it is no longer necessary to perform an aseptic technique. Use soap and water to clean the stoma site and ensure the area around the stoma is dried thoroughly. The tube tip should be cleaned daily using water and a small brush (Loser et al. [90]).
Where the enteral feeding tube has been placed to allow nutritional support, the patient's feeding regimen should be checked. If there is no feeding regimen, liaise with dietetic staff as needed or contact the medical team.
Ongoing care
To reduce the risk of tube blockage, the patient's medication should be reviewed where the tube is required for drug administration. Liaise with both the medical team and the pharmacist to discuss the patient's prescription and alternative routes and medications (NICE and BNF [125]).
Education of the patient and relevant others
The need for patient and/or carer teaching should be considered in patients with newly inserted enteral feeding tubes, and discharge planning should be initiated. If a patient is discharged within 72 hours of gastrostomy insertion, the patient and their carers must be provided with appropriate local and out‐of‐hours contacts for urgent aftercare advice and warned of the danger signs that need urgent attention. These include pain on feeding, new bleeding and external leakage of gastric contents (NHSI [116]).
Complications
Infection
Infection can occur at the time of placement or be due to inappropriate site care, particularly prior to tract maturation. Patients who are immunocompromised may be at increased risk of infection. Initial care after insertion should be carried out using an aseptic technique until the tract is fully mature. If there are any signs of reddened skin or infection then a swab should be taken and sent off for microbiology culture and sensitivity. Appropriate antibiotic prescribing may be required depending on the condition of the stoma site and the microbiology results from the swab (Warriner and Spruce [185]).
Pressure necrosis
Excessive tightening of the external fixator device may cause pressure necrosis around the stoma site. The external fixator should be adjusted while the patient is sitting up, where possible. If the external fixator is adjusted while the patient is lying down then it is necessary to ensure that it is applied 1–2 cm from the skin surface.
The stoma site should be reviewed regularly and the patient and carer taught how to adjust and reposition the external fixator. This is particularly important if the patient has gained weight as this may cause pressure on the fixator.
Buried bumper
In enteral feeding tubes that have a bumper as the retention device, the retention bumper may become buried in the intestinal wall. The external fixator should be released and the tube pushed in approximately 5 cm and rotated fully both clockwise and anticlockwise. Ensure it is rotated regularly according to the manufacturer's instructions (NNNG [127]).
Peristomal leakage
It is important to examine the tube site for causes of peristomal leakage. Contributory factors include delayed gastric emptying, buried bumper and infection. Ensure correct management of the stoma area, including checking the internal balloon weekly if this is the method of tube retention. If there is delayed gastric emptying, investigate and treat the cause of this. Avoid use of bulky dressings under the external fixator. Consider skin care and use protective cream or spray if necessary. Liaise with medical colleagues and consider the need for antacid medication to reduce skin excoriation (Westaby et al. [188]).
Pneumoperitoneum
This is a known complication of the insertion procedure when air is present in the abdominal cavity. The patient's medical team must be contacted immediately if the patient is complaining of increased pain or pyrexia or has signs of an acute abdomen.
Unplanned removal of tube
A gastrostomy or jejunostomy that is accidentally removed within 14 days of insertion requires immediate attention (Westaby et al. [188]). It should be replaced endoscopically or radiologically as soon as possible as a mature tract will not have developed in this period (Lynch and Fang [92]).
The patient's medical team or nutrition nurse must be contacted immediately, as the stoma site can heal over within 2 hours. The leakage of gastric contents into an immature tract carries the risk of infection. See the section on enteral silicone plugs (below) for advice on how to minimize the risk of stoma occlusion.
Hypergranulation
There is a lack of high‐quality evidence in the management of hypergranulation and so the treatment of this condition can be difficult. It is recognized as a common problem but classed as minor in the literature and so there is no standard practice of care (Warriner and Spruce [185]).
Excess movement of the enteral feeding tube within the tract may stimulate the growth of granulation tissue. Ensure that the fixation plate is firm but not too tight on the abdomen to minimize movement of the tube. If the tube does not have an external fixator plate then consider placing one to stop excessive movement of the tube (NNNG [127]). Other factors that may influence the formation of hypergranulation include excess moisture, critical colonization or true infection, and the presence of foreign material (NNNG [127]).
Local guidance on the management of hypergranulation should be agreed and be based on current evidence and the patient population.
Tube inspection
The tube should also be inspected for damage, such as splitting or cracking, during routine daily hygiene. Any damage to the tube could enable a leaking of gastric content onto the surrounding skin, resulting in skin breakdown or excoriation. Adherence to these principles is thought to contribute to the prevention or early detection of problems associated with enterostomal feeding.
Silicone gastrostomy stoma plugs
Definition
Silicone gastrostomy stoma plugs are designed for emergency use to keep stomas from closing when a gastrostomy device has fallen out. They are a temporary measure to minimize the risk of the stoma closing completely and come in several lengths and gauges to fit the need of the patient.
Procedure guideline 8.17
Insertion of a silicone gastrostomy stoma plug
Administration of enteral tube feed
Pre‐procedural considerations
Prior to using enteral feeding tubes for medication or feed administration, it is vital to know where in the gastrointestinal tract the tube tip lies. This may be difficult in patients who have tubes placed in other organs or where there is little visible difference externally between gastrostomy, transgastric or jejunostomy tubes. Where possible, the tube size, type, insertion date and method should be clearly documented. If this information is unavailable, the tube should be aspirated and the pH used to differentiate between gastric and small bowel placement. The pH in the jejunum is usually between 7 and 9 (neutral or slightly alkaline) whereas the pH of the stomach is below 5.5. If there are sutures securing the external fixator to the patient's abdomen, these should not be removed until it has been confirmed that they are not required to keep the tube in position.
Procedure guideline 8.18
Enteral feeding tubes: administration of feed using an enteral feeding pump
Table 8.16 Prevention and resolution (Procedure guideline 8.18)
Problem | Cause | Prevention | Action |
---|---|---|---|
Pump alarm states ‘occlusion’ or ‘empty’ | The feed may have finished. There may be a blockage in the giving set or feeding tube. | Ensure the feed container was shaken well before feeding. Ensure the giving set was not kinked when feeding was commenced. Ensure that the roller clamp/tap is fully open. Flush the feeding tube as directed before commencing. | Straighten any kinks in the giving set. Ensure that the giving set is fixed correctly around the rotor. Open the roller clamp/tap fully. Check that the feeding tube is not blocked. Disconnect from the feeding tube and run the feed into a container; if feed runs and there is no alarm, this indicates that the pump is working properly and the feeding tube is probably blocked. |
Pump alarm states ‘low battery’ | This indicates that the pump battery needs to be recharged. | Keep pump plugged in and charged. | Connect to the mains power and continue to feed. |
Unable to prime giving set | The roller clamp/tap may not be fully open. There may be a fault with the giving set. If a drip chamber is present then feed may not have run into this. Also check the pump is functioning correctly to initiate the prime. | Ensure that the roller clamp/tap is fully open when beginning to prime the giving set. | Open the roller clamp/tap fully. Squeeze some feed into the drip chamber if applicable. Try with a new giving set. |
Post‐procedural considerations
Immediate care
As soon as the feed commences, check that it appears to be running without problems and at the correct rate. Monitor this regularly throughout the feed administration.
Monitor the patient for signs of nausea or abdominal discomfort within the first hour and every 2–4 hours during feed administration. This may not be possible if the feed is given overnight and the patient is asleep.
Ongoing care
In order to avoid complications and ensure optimal nutritional status, it is important to monitor the following in patients requiring enteral tube feeding:
- oral intake
- bodyweight
- urea and electrolytes
- blood glucose
- full blood count
- fluid balance
- tolerance to feed, for example nausea, fullness and bowel activity
- quantity of feed taken
- care of tube
- care of stoma site (where appropriate).
If appropriate, the patient should be taught how to follow the procedure of setting up the enteral feeding equipment. They should be confident with the maintenance of the equipment and be aware of how to troubleshoot.
Complications
Aspiration
This may occur due to regurgitation of feed, poor gastric emptying or incorrect placement of a nasogastric tube. The risk of this can be reduced by:
- the use of prokinetics (e.g. metoclopramide), which encourage gastric emptying
- checking the position of the tube before feeding
- ensuring the patient has their head at a 45° angle during feeding; if the patient is in bed then this can be achieved through raising the head of the bed and ensuring the patient has sufficient pillows for support (Gandy [58]).
Nausea and vomiting
This can be caused by a number of factors. It can be related to disease, a side‐effect of treatment or a medication such as antibiotics or analgesia. A combination of poor gastric emptying and rapid infusion rates can also stimulate nausea and vomiting. Nausea and vomiting can best be controlled through the use of antiemetics, a reduction in the infusion rate or a change from bolus to intermittent feeding.
Diarrhoea
This could be a result of:
- medications such as antibiotics, chemotherapy or laxatives
- disease or treatment, for example pancreatic insufficiency or bile acid malabsorption
- gut infection, for example Clostridioides difficile
- poor tolerance to the feed.
Antidiarrhoeal agents can be used if a person is experiencing diarrhoea as a side effect of medication. If possible, an alternative medication should be found that does not cause diarrhoea. When the diarrhoea is disease related, the underlying problem should be treated; that is, if a person has pancreatic insufficiency, they should be provided with a pancreatic enzyme supplement and/or peptide‐based feed.
Avoiding microbiological contamination of the feed and equipment will help to reduce the risk of diarrhoea. This will involve keeping the equipment clean and, when feeding, maintaining a sealed system.
A stool sample should be sent to microbiology to test for cultures and sensitivities to check for any gut infection. If the sample is found to be positive then the infection should be treated appropriately. Refer to Chapter c06: Elimination for further information on the diagnosis and treatment of diarrhoea.
If all of the above have been ruled out, the dietitian can review the osmolarity, fibre content and infusion rate (Todorovic and Mafrici [181]).
Constipation
Constipation can be caused by inadequate fluid intake, immobility, bowel obstruction, or the use of opiates or other medications causing gut stasis. Methods to improve symptoms of constipation include:
- checking fluid balance and increasing fluid intake if necessary
- providing laxatives or bulking agents
- if possible, encouraging mobility
- if there is a bowel obstruction, discontinuing enteral feeding (Todorovic and Mafrici [181]).
Abdominal distension
This can be caused by poor gastric emptying, rapid infusion of feed, constipation or diarrhoea. Possible ways to improve distension include:
- gastric motility agents
- reducing the rate of infusion
- encouraging mobility if possible
- treating constipation or diarrhoea.
Blocked tube
Blockage can be a result of inadequate flushing, failure to flush the feeding tube or administration of inappropriate medications via the tube. For ways to unblock a feeding tube refer to Procedure guideline 8.20: Enteral feeding tubes: unblocking.
Enteral feeding tubes: administration of medication
Evidence‐based approaches
Rationale
Indications
Use of an enteral feeding tube for the administration of medication is indicated for patients requiring medications who are not able to take oral preparations due to dysphagia. Where possible, medications should be administered in liquid or soluble form. Alternatively, some preparations can be given in sublingual form.
Contraindications
There are no absolute contraindications to the use of enteral feeding tubes. However:
Other considerations when deciding whether to use the tube for medication administration are as follows:
- Do not instil any medication that poses a risk of blocking the feeding tube.
- Enteric coated or slow-release medications should not be crushed and are therefore not suitable for administration down an enteral tube.
- If the preparation would be harmful to the administrator, seek advice from a pharmacist.
- It is not advisable to mix different medications prior to administration unless advised to do so by a pharmacist.
- 60 mL ENFit syringes must not be used to measure drugs where the required drug volume is below 10 mL as they are not sufficiently accurate at such volumes. A 10 mL oral syringe must be used to measure drug volumes and the dose then transferred to an appropriate ENFit syringe for administration. Drug volumes of 10 mL and above can be measured using the ENFit syringe that is going to be used to administer the drug.
Procedure guideline 8.19
Enteral feeding tubes: administration of medication
Table 8.17 Prevention and resolution (Procedure guideline 8.19)
Problem | Cause | Prevention | Action |
---|---|---|---|
Tube blocked with medication | Medication was not administered in the correct composition and/or the tube was not flushed adequately | Ensure that the guidance from the pharmacist is followed correctly. Ensure that the tube is flushed before and after administration. | See Procedure guideline 8.20: Enteral feeding tubes: unblocking. |
Unable to administer required medication | Medication is in a form that cannot be crushed or dissolved | Ensure that the pharmacist and medical team are aware that all medications need to be administered via an enteral tube. | Contact the medical team or pharmacist to seek advice on an alternative medication. |
Post‐procedural considerations
Immediate care
The tube's patency should be checked to ensure that the medication has not caused a blockage. This can be done by flushing the tube. The patient should be monitored to ensure that there are no unanticipated side‐effects of the medication administered.
Education of the patient and relevant others
If the patient is going home with the enteral tube in place, it should be ensured that the patient is educated and confident with administering their feed and medication. If this is not possible then the patient should be referred to a healthcare professional, such as a community nurse, who can undertake this aspect of care.
Home enteral feeding
Some patients who are established on tube feeding in hospital also require enteral tube feeding at home. A multidisciplinary approach is needed for a successful discharge, usually involving a dietician, doctor, ward nurse, community nurse and GP. The patient's circumstances and the ability of the patient or carers to manage the feed must be considered when discharge is being planned. Adequate time should be allowed in the hospital setting for patients to become fully accustomed to the techniques of feed administration and care of the feeding tube, prior to discharge home. Patients should also be given written information to reinforce the education they receive prior to discharge (BAPEN [11]).
Support in the form of the GP, community nurse and community dietetic services should be established before discharge. A multidisciplinary discharge meeting may be of benefit to both the patient and the professionals involved. Many of the commercial feed companies can organize for a patient's feed and equipment to be delivered to their home, after consultation with the local community services (BAPEN [11]). The hospital or community dietician can arrange this. Early notification of discharge is essential to facilitate this transition of care.
Termination of enteral tube feeding
It is important to ensure that an individual is able to meet their nutritional requirements orally prior to termination of the feed. Enteral tube feeding may be discontinued when oral intake is established (NCCAC [108]). It may be useful to maintain an overnight feed while the patient is establishing oral intake (Pearce and Duncan [138]).
Procedure guideline 8.20
Enteral feeding tubes: unblocking
Table 8.18 Prevention and resolution (Procedure guideline 8.20)
Problem | Cause | Prevention | Action |
---|---|---|---|
Cracked/split tube | Too much pressure applied while flushing tube | Avoid applying too much pressure while flushing the tube and take care with a syringe smaller than 50 mL. | If the tube cracks or splits then it must be replaced. |