Pre‐procedural considerations

Equipment

12‐lead ECG machine

A 12‐lead ECG machine has 10 cables, which are connected to electrodes fixed to the body. The machine detects and amplifies the electrical impulses that occur at each heart beat and record the waveforms onto graphed paper (using a stylus) or onto a computer. All ECG machines should be tested to ensure accurate data are recorded. Calibration is usually undertaken by qualified engineers; however, this should be verified in the clinical setting to confirm that when a voltage of (usually) 1 mV over 0.2 seconds is put through the machine, the machine produces the expected deflection of 10 mm and the paper moves at the correct speed: 25 mm per second (Aehlert [3], SCST [183]). This calibration verification should be printed at the beginning or end of each line of the ECG as a box shape with a 5 mm width and a 10 mm height (Figure 14.17) (Wesley [209]).
image
Figure 14.17  Electrocardiogram (ECG) calibration verification signal. Source: Reproduced from Crawford and Doherty ([49]) with permission of MA Healthcare.

Electrodes

The 10 electrodes are attached to the patient's body in the positions described above and in Procedure guideline 14.2: Electrocardiogram (ECG). The cables clip onto these electrodes and record 12 views of the heart using different combinations of electrodes to measure various signals from the heart.

ECG paper

ECG graph paper is divided into small squares of 1 mm each, and five small squares make up a large square. The horizontal axis of the ECG paper represents time and the vertical axis represents amplitude. Standard ECG paper speed is 25 mm per second; each small square equals 0.04 seconds and one large square equals 0.2 seconds, so five large squares equal 1 second. ECG machines are calibrated so that the deflection amplitude of cardiac conduction is measured in mm/mV; 10 small squares or two large squares show a deflection of 1 mV (10 mm/mV) (see Figure 14.17) (Aehlert [3], Wesley [209]).

Filter

Filters are electronic devices that remove artefacts from the ECG, improving the tracing obtained. ECG machines have internal pre‐programmed filters set by the manufacturer or clinical engineers, which cannot be altered by clinical staff, and also control panel filters, which can be deployed by clinical staff. However, these can distort the ECG waveform, and therefore are not recommended for routine use, but they may be required if all other actions to reduce artefacts have been unsuccessful. If a filter is used, this must be clearly documented on the ECG (SCST [183], Wesley [209]).

Maintenance and storage

All ECG machines should be used and maintained in accordance with the manufacturer's instructions. In addition, as they are often used in emergency situations, they should always be available and in good working order (SCST [183]). After use, the ECG machine should be cleaned, returned to its storage location and plugged into a mains electrical supply to charge the internal battery ready for use in an emergency (SCST [183]). Any malfunctioning machine should be taken out of clinical use until repair or service can be undertaken by the relevant department, in accordance with local policy (Wesley [209]).

Specific patient preparation

Female patients

The conventional placement of the lateral chest leads (V4, V5 and V6) is beneath the left breast. While there is emerging evidence to support the positioning of these over the breast, it is insufficient to suggest a change of procedure (SCST [183]). It is also worth noting that the fifth intercostal space can only be found by lifting the breast and therefore it is logical to position the electrodes here (SCST [183]).

Dextrocardia

Dextrocardia is any situation where the heart is located within the right side of the chest rather than the left. This may be associated with the condition situs inversus, believed to occur in 1 out of 8000 people, where all of the patient's organs are in mirror‐image positions (Sharma et al. [186]).
The SCST ([183]) suggests the following approach to ECG recording for patients with suspected or known dextrocardia:
  • Dextrocardia should be suspected if the ECG shows an inverted P wave in lead I (P‐axis >90°) together with poor R wave progression across the chest leads. In this case a second ECG should be recorded with the chest electrodes positioned on the right side of the chest using the same intercostal spacing and anatomical landmarks to provide a true ECG representation (Figure 14.18). The limb lead complexes will continue to appear inverted, demonstrating the abnormal location of the heart, but the repositioned chest leads will now show the appropriate R wave progression. The revised electrode positions should be clearly documented on the second ECG, and both ECGs should be retained for inclusion in the patient's notes.
  • Patients who are known to have dextrocardia should have the ECG recorded with the limb electrodes in the usual positions and the chest electrodes placed across the right side of the chest, as described above. Note that swapping of the right‐ and left‐limb electrodes will normalize the appearance of the limb leads; therefore, when repositioning electrodes, it is imperative that the ECG is clearly annotated to describe the new positions of the electrodes (V3R, V4R, etc.) to prevent the possibility of dextrocardia being overlooked.
image
Figure 14.18  Right‐sided chest lead positioning for an electrocardiogram (ECG).

Education

The patient must give their consent for the ECG to be performed, after appropriate explanation of what the procedure consists of, the rationale and the intent (NMC [144], Peate and Wild [157]). The patient's privacy and dignity should be ensured and reassurance should be provided, along with an explanation that an ECG is not a painful procedure (NMC [144], SCST [183]). The patient should be positioned comfortably either lying or sitting, preferably in a semi‐recumbent position (at an angle of 45°) with their head supported (SCST [183], Wesley [209]). It is important that the patient is relaxed and keeps still during the procedure to reduce interference and artefacts and to facilitate recording of a clear and stable ECG trace (Aehlert [3], SCST [183]).
Procedure guideline 14.2
Table 14.4  Prevention and resolution (Procedure guideline 14.2)
ProblemCausePreventionAction
Unable to turn on the ECG machineLow batteryEnsure the ECG is left on continuous charge when not in use.Connect the ECG machine to mains electricity. It should function once it has been connected to a power supply.
The ECG machine is working but the rhythm display is blankLoose connectionCarefully store the cables after use to prevent damage.Check that each lead is connected to the electrode clip and that the base of the cable is connected to the ECG machine.
Electrode stickers peel off from the patient's skin
Prepare the patient's skin prior to electrode application (see step 4 of Procedure guideline 14.2: Electrocardiogram) (ECG).
Cleanse the electrode sites with soap and water to remove any lotion or reduce skin moisture, and allow to dry. Reapply new electrode stickers.
If the patient has a lot of chest hair, try to push the hair out of the way when applying stickers to make a better contact; reinforcement with tape may be required. If this is unsuccessful, ask the patient for permission to clip or shave the hair at the electrode sites, cleanse to remove loose hair, allow to dry and apply new electrode stickers.
The electrode stickers may be intact but there is still no rhythm – the patient's skin may be excessively dry and flakyIdentify underlying causes and severity of dry skin and address them with the multidisciplinary team as appropriate.With the patient's consent, perform vigorous but gentle rubbing with soap and water to remove the superficial dead skin cells. The skin should be allowed to dry and new electrode stickers applied.
The ECG is printing but some of the views are missingA loose connectionEnsure good skin preparation and maintenance of the ECG machine.Check that the electrode stickers are intact and follow the above suggestions for skin preparation. Check that the cables are all securely connected.
The ECG is printing but there is interference, making it difficult to interpret accuratelyPatient movementAsk the patient to remain still and not to speak, but to breathe normally.Ensure the patient is not moving or talking and repeat the ECG once the displayed rhythm has stabilized.
The patient is cold and therefore involuntarily shiveringIf at all possible and not clinically dangerous, wait for the patient to warm up.Once the electrodes are connected, place a gown, sheet or blanket over the patient to warm them.
Peripheral interferenceAs above – ask the patient to remain still and ensure they are warm.Place the limb electrodes more centrally, as shown in the Mason–Likar system (see Figure 14.16), to reduce interference from limb movement.
General interference or a wandering baselineEnsure there is good skin contact with the electrodes and good electrode placement, and ask the patient to avoid thoracic movement.
Ensure good skin preparation.
Ask the patient to lie still and breathe normally.
Check that the electrodes are properly placed.
If necessary, record an additional ECG with the ‘filter’ function on to limit interference. Write on the ECG that the filter function was used.
Electrical interference from nearby devices (such as infusion pumps, other monitoring devices or haemodialysis machines)It may be necessary to temporarily remove or suspend electrical devices if safe and appropriate to do so.
Repeat the ECG once the displayed rhythm has stabilized.
If necessary, record an additional ECG with the ‘filter’ function on to limit interference. Write on the ECG that the filter function was used.
The ECG is working and displays a rhythm but the print‐out is blankA different manufacturer's ECG paper has been loadedOnly use the manufacturer's recommended paper for the ECG machine.Reload the machine using the correct paper and repeat the recording.
Internal faultEnsure the manufacturer's recommendations on maintenance and servicing are followed.Contact the in‐house engineer, medical device technician or manufacturer for advice in accordance with local policy.